A hydrogen peroxide biosensor based on the bioelectrocatalysis of hemoglobin incorporated in a kieselgubr film

Haiyan Wang,Rong Guan,Chunhai Fan,Dexu Zhu,Genxi Li
DOI: https://doi.org/10.1016/S0925-4005(02)00027-8
2002-01-01
Abstract:An unmediated hydrogen peroxide biosensor was prepared by co-immobilizing hemoglobin (Hb) with kieselgubr at a pyrolytic graphite electrode surface. Hb exhibited direct electrochemistry in the film, as well as peroxidase-like activity. This modified electrode showed high stability, catalytic activity towards the reduction of hydrogen peroxide. A linear dependence of the catalytic current on the concentration of hydrogen peroxide was observed in the concentration range of 5.0×10−6 to 3.0×10−4mol/l with a detection limit of 2.1×10−6mol/l. The apparent Michaelis-Menten constant (KMapp) was estimated to be 975μmol/l.
What problem does this paper attempt to address?