NEW METHOD FOR MEASUREMENT OF DEFORMABILITY OF RBCs USING MICROCHANNELS FORMED IN A SILICON SUBSTRATE

Xiaolin Zheng,Nan Wu,Xiao Lu,Shaoxi Cai,Yunpeng Wu,Chenglin Peng,Lin He,Ganyi Hu,Wo Zheng,Maojing Cui
DOI: https://doi.org/10.3969/j.issn.0258-8021.2000.01.002
2000-01-01
Abstract:Method and technique of the simulation system of capillary were studied by microchannels formed in a silicon substrate. The design principle and the method were based on the micromechanical technology. The paper presented the manufacturing process of silicon microchannels. The channel grooves were measured at submicrometer accuracy, and formed by the technology of photolithography and etching for integrated circuits. The full process of the flow of cells through the microchannels can be quantitatively observed. The author also studied the reproducibility and application experiment of the method. The experimental results proved that the a new method has a good reproducibility (Coefficient of variation CV < 5%). The new method was is informative, accurate, convenient and economical.
What problem does this paper attempt to address?