Pt-Ir alloy stent made in China implanted in the coronary artery of dog

朱建国,崔长琮,兰燕平,李满祥
DOI: https://doi.org/10.3969/j.issn.1671-8259.2000.03.005
2000-01-01
Abstract:Objective: To assess the biocopatibility and the relevant physics characteristics of domestic Platinum-Iridium alloy stent. Methods: The domestic stents were implanted in the left circumflex (LCX) of coronary artery of 16 dogs. The histologic examination was carried on in 1 week (n = 5), 4 weeks (n = 5) and 24 weeks (n = 5) respectively. Results: The stents were successfully implanted into the middle segment of LCX in 16 animals. One died of acute thrombosis within 21 hours of stent implantation . Angiography revealed all vessels were patent; Histopathologic examination revealed the neointimal tissue began to cover the stent wire at 1 week ,few inflammatory cells were seen near the stent struts. At 4 weeks , the stent struts were embeded in the artery tissue, almost no inflammatory cells were seen. In 24 weeks ,neointimal tissue was thinner than that of 4 weeks, no inflammatory cell was extravasated. Conclusion: This study demonstrates that the domestic Pt-Ir alloy stent ,which is simply using, easy flexibility to cross the curving coronary artery, high radial force, good biocompatibility and vessel patence, accords with international stent standard.
What problem does this paper attempt to address?