Experimental Novel Absorbable Magnesium Alloy Stent

杨水祥,雷力成,王丽丽,昌红,王尔德,吴坤,崔福斋,杨静罄,朱锦文
DOI: https://doi.org/10.3969/j.issn.1009-0126.2012.07.022
2012-01-01
Abstract:Objective To assess the biocompatibility,efficacy and safty of a self designed and manifactured novel absorbable magnesium alloy stent.Methods An absorbable magnesium alloy stent was implanted into the coronary and femoral artery of 35 hybrid dogs,respectively.The implanted stents were observed by coronary arteriography and femoral arteriography,respectively.The isolated blood vessels in the stent segment were observed by histopathology and their computer images were analyzed on days 1,3,5,7,14,21 and 28 after operation,and five dogs in each group.Results Fifty-one stents were successfully implanted into the coronary and femoral artery of 35 dogs.Coronary and femoral arteriography showed that the stenting vessels were patent with no stenosis and thrombosis.Histopathology showed no intimal hyperplasia but residual stent on days 1,3 and 5 after operation.The stent was completely degraded on day 7 after operation.The intimal hyperplasia was observed on day 14 after operation and became significant on days 21—28.No significant inflammatory reaction and thrombosis were found at different time points after operation. The intimal hyperplasia area was 0.04±0.03 mm~2,0.10±0.03 mm~2 and 0.15±0.04 mm~2,respectively on days 14,21 and 28 after operation,accounting for 1.84%±1.18%,3.72%±1.12% and 6.29%±3.36%,respectively(P0.05).Conclusion Absorbable magnesium alloy stent is easy to manipulate with a good biocompatibility,efficacy and safety,and is thus of a vast prospect in its clinical application.