Abstract:AIM:To evaluate the biocompatibility of plasma coated stents in normal canine coronary arteries.METHODS:Implation of plasma coated and non-coated stents was performed in the canine coronary arteries. The animals were killed after 3 d(n=4),15 d(n=4),1 month(n=5),and the vessels were harvested for scanning electron microscopy,histology,and morphometry.RESULTS:Stent thrombosis did not occur in either group and platelet adhesion were similar in both group at 3 d after stenting. At 15 d both the coated and non-coated stents were equally endothelialised. At 1 month there were no difference in intima thickness between the coated and non-coated stents and did not appear to elicit an adverse inflammatory response. CONCLUSION:In this animal model the plasma coating showed excellent blood and tissue compatibility.