Group-phase Refractive Index Method for Improving the Accuracy in Two-Color Interferometric Length Measurements

LJ Zeng,H Matsumoto,K Seta
DOI: https://doi.org/10.1063/1.1149850
IF: 1.6
1999-01-01
Review of Scientific Instruments
Abstract:A group-phase refractive index method was developed for improving the accuracy of a conventional two-color method used in an interferometer for length measurements. This method uses two closer wavelengths to reduce the errors caused by air turbulence and the chromatic aberration. We describe the new method and demonstrate preliminarily its effectiveness by measuring the group refractive index of air using a vacuum cell. We then compared the results with those calculated using the Edlén equation. The mean difference between the measured and calculated values was 2.15×10−6 with a standard deviation of 2.0×10−6 for the vacuum cell with a length of 184.5 mm.
What problem does this paper attempt to address?