Investigation on radioactive rays ionization gas flowmeter

Yikun Qu,LiYe Zhou,Qitian Miao,Renzhi Wei
1997-01-01
Abstract:A microcomputer controlling gas flow velocity measuring system based on the detection of the gas ionized by �� rays is presented. A detector with the ion collecting plate parallel to the radioactive rays has proven effective in the collection of ions. The ion cloud outline modulated by a step motor has been described with its size calculation formula presented; the basic control relation for the main parameters to maintain high measuring precision has been proposed; the necessary voltage needed for the effective collection of the ions has been derived. It is due to the configuration of the detecting plates that the precision of this ion detection gas flow device is around 2% and is reliably resistant to dust and moisture after several days operation in laboratory. The timing error caused by amplitude fluctuation is diminished by the peak searching signal processing method utilized.
What problem does this paper attempt to address?