Abstract:Based on the principle that the position and intensity of Michelson interference fringes change due to different optical path lengths in different gases, a new system is designed, which applies two interferometers, one is used for test and the other is used as reference. The problems of low precision and bad reliability existing in detecting gas concentration are solved. In this system UNIQ UM201 Charge Coupled Device and Matrox Meteor-II image grab card are used to acquire the interference fringes and the images of the obtained fringes are processed in the LabVIEW environment. As the surroundings change faintly, the images will also change, and this test scheme has high sensitivity and fast response.