Application of KCl-denuder for measuring gas-phase mercuric species (Hg2+) in atmosphere

Shiqiang Wei,Lindqvist Oliver,Sommar Jonas
IF: 6.9
1997-01-01
Journal of Environmental Sciences
Abstract:KCl denuder technique was employed in the measurement of Hg(II) species in atmosphere firstly. Intensive lab tests indicate that Hg0 can pass through KCl denuder freely, while Hg2+ can be trapped in it very efficiently. The efficiencies for adsorbing inorganic Hg2+ averaged between 95% to 100% with a gross mean of 98%, and those for absorption of organic Hg2+ were around 94%. Mass balance calculation shows that mercury trapped in denuder can be quantitatively extracted by 1 mol/L HCl (super grade) and analysed by the method of SnCl2 reduction-CVAFS determination, thus, demonstrated the feasibility of KCl denuder in measurement of Hg(II) in atmosphere. Preliminary application at two sites,G?teborg, Sweden and one site in Ireland, found 0.04-0.15 ng/m3 of Hg(II) species, contributing to 2%-10% of the total mercury.
What problem does this paper attempt to address?