Determination of afterload in simulation blood circulation system

Zhaorong Liu,Guanghong Ding,Xixi Li
1994-01-01
Abstract:In almost all of the duplicator for pulsatile flow testing of heart valve substitute (or the simulation device of the blood circulation system), there are many problems such as, repeat the physiological condition is not well, the objectivity and the comparability of the experimental result are very poor. In this paper, we used arterial input impedance as a parameter of ventricular afterload instead of tradional use of waveforms of pressure and flow, in order to overcome disadvantages of comparability and limitation of physiological condition. An afterload system of blood circulation simulation was deviced with better comparability and imitation. | In almost all of the duplicators for pulsatile flow testing of heart valve substitute (or the simulation devices of the blood circulation system), there are many problems such as, repeating the physiological condition is not good, the objectivity and the comparability of the experimental results are very poor. In this paper, arterial input impedance was used as a parameter of ventricular afterload instead of traditional use of waveforms of pressure and flow, in order to overcome disadvantages of comparability and limitation of physiological condition. An afterload system of blood circulation simulation was developed with better comparability and limitation.