Multi-Modal in Vitro Experiments Mimicking the Flow Through a Mitral Heart Valve Phantom
Lea Christierson,Petter Frieberg,Tania Lala,Johannes Töger,Petru Liuba,Johan Revstedt,Hanna Isaksson,Nina Hakacova
DOI: https://doi.org/10.1007/s13239-024-00732-3
2024-05-24
Cardiovascular Engineering and Technology
Abstract:Fluid-structure interaction (FSI) models are more commonly applied in medical research as computational power is increasing. However, understanding the accuracy of FSI models is crucial, especially in the context of heart valve disease in patient-specific models. Therefore, this study aimed to create a multi-modal benchmarking data set for cardiac-inspired FSI models, based on clinically important parameters, such as the pressure, velocity, and valve opening, with an in vitro phantom setup.
engineering, biomedical,cardiac & cardiovascular systems