Abstract:The dynamic aortic valve (DAV) is a new left ventricular assist device, a micro-axial blood pump implemented at the position of the aortic valve, pumping blood from the left ventricle into the aortic artery. The present dynamic aortic valve operates at 7 different rotation speeds, ranging from 3000 r/min (speed 1) to 9000 r/min (speed 7). Because in vivo experiments need a lot of live animals and take a long period of time, modeling and simulation have been widely used to simulate and analyze hydra-dynamic property of the DAV and its assisting effects. With the measurements from the mock circulatory loop, a mathematic model of the DAV is established and embedded into the previously developed canine circulatory system. Using this model, the effect of the DAV on the failing heart at each rotation speed level is investigated. The vital cardiac variables are computed and compared with in vivo experimental results, which are in good agreement with an acceptable difference mostly 15%. The establishment of the DAV model and its simulation are useful for further improvement of the DAV device.
What problem does this paper attempt to address?