A Simulation Study on Cardiac Assist Device: Hemopump

李熊,白净,何平
DOI: https://doi.org/10.3969/j.issn.1002-0837.2003.01.013
2003-01-01
Abstract:Objective To study the interaction between HP31 Hemopump system and the cardiovascular system. Method A dynamic model for a HP31 Hemopump system was developed basing on theoretical analysis and experimental data. By incorporating the Hemopump model in a previously established model for the canine circulatory system, the effects of the pump on various hemodynamic variables of the circulatory system were investigated and the benefit of the pump to a failing heart was demonstrated. In addition, different mechanisms of the Hemopump and IABP that relieve the workload of left ventricle were compared. Result Our results verified that the Hemopump assists a failing heart by increasing the oxygen supply to the heart while reducing its oxygen consumption through a reduction in the preload of left ventricle. These beneficial effects are generally enhanced when the pump′s rotation speed is increased. However, when the pump rotation speed is too high, the inflow to the pump may be impaired and the pump performance will be negatively affected. Conclusion Predications from the model are in good agreement with the results previously obtained in animal experiment and in vitro measurements. The model described in this paper should provide a useful tool to investigate and test various strategies for optimal control of the Hemopump.