Hemolysis and Animal Experiments with the Magnetically Drived Axial Flow Ventricular Assist Device

吴广辉,蔺嫦燕,侯晓彤,张锡文,朱来来,李冰一,王景
DOI: https://doi.org/10.3969/j.issn.1009-7090.2010.01.001
2010-01-01
Abstract:Objective To design and develop a magnetically drived axial left ventricular assist device,and evaluate its performance. Methods For the test of hemolysis in vitro mock circuit,a particular blood bag was used as a mock circulation loop and the sheep blood were used as the circulating medium. The normalized index of hemoglobin (NIH) was used to measure the hemolysis. The in vivo adaptability was tested through 12 hours in animal experiments with 3 goats. Results The NIH for the axial blood pump was (0.158 ± 0.043) mg/L. There were no mechanical breakdowns in 12 hours animal experiments; the free hemoglobin (FHb) of each experimental animal started rising just after circulation; the highest value was 164.8 mg/L. Conclusion The results obtained from the in vivo testing and animal experiment indicated that this is an acceptable design for blood pump,and worth to further improvement.