Sensitivity compensation method in ultrasonic inspection for curved components with variable thickness

张杨,周晓军,杨辰龙,李雄兵,郑慧峰
DOI: https://doi.org/10.3785/j.issn.1008-973X.2013.01.017
2013-01-01
Abstract:A new compensation method was proposed in order to solve the compensation problem in ultrasonic inspection for curved components with variable thickness. The method combined sound field calculation and experimental measurement technique. The Gaussian beam was introduced in order to calculate the receiving transducer ultrasonic field. The analytical expressions for the amplitude and phase of a Gaussian beam were described after propagation through media and interactions with curved interfaces. According to the conditions such as component thickness, shape and probe location, the receiving sound model of through-transmission ultrasonic inspection system was deduced based on multi-Gaussian beam superposition method. In practice, additional attenuation caused by medium absorption and scattering should be considered. After the attenuation coefficient was measured by the method named inserted and substituted technique, the receiving sound model was modified. Then the receiving sound pressure was predicted with different parameters based on the modified model, and attenuation of the receiving sound pressure was calculated for sensitivity compensation. Experimental results showed the correctness and effectiveness of the new method.
What problem does this paper attempt to address?