A MEMS Flow Sensor for Self-adjusted Precise Non-Contact Liquid Dispensing

Liu Yaxin,Zhong Ming,Yao Yufeng
DOI: https://doi.org/10.19026/rjaset.5.5047
2013-01-01
Abstract:A MEMS flow sensor was proposed to enhance the reliability and accuracy of liquid dispensing system.Benefiting from the feedback of sensor information, the system can self-adjust the open time of the solenoid vale to accurately dispensing desired reagent volume without pre-calibration.This study focuses on the design, fabrication and application of this flow sensor.Firstly, the design, fabrication and characteristics of the MEMS flow sensor based on the measurement of the pressure difference across a flow channel are presented.Secondly, the liquid dispensing system in which the flow sensor is integrated will be introduced.A novel closed-loop control strategy is proposed to calculate valve open-time for each dispensing cycle.Finally, experiments results are presented with different dispensing volumes, Coefficient of Variance (CV) has been shown to be below 3%.It indicates that integration of the MEMS flow sensor and using of a compound intelligent control strategy make the system immune to liquid viscosity, pressure fluctuation and some other disturbances.
What problem does this paper attempt to address?