Comparison of sorbents for the solid-phase extraction of the highly polar degradation products of atrazine (including ammeline, ammelide and cyanuric acid)
V PICHON,L CHEN,S GUENU,M HENNION
DOI: https://doi.org/10.1016/0021-9673(95)00527-T
IF: 4.601
1995-01-01
Journal of Chromatography A
Abstract:Three different types of non-polar sorbents, i.e., C18 silica, apolar styrene-divinylbenzene copolymer (PRP-1) and porous graphitic carbon (PGC), were compared for the solid-phase extraction of atrazine, simazine and nine degradation products obtained by dealkylation, dehalogenation, hydroxylation or deamination mechanisms. The C18 silica was shown to be effective only for the first degradation products (deethyl-, deisopropyl- and hydroxyatrazine), the others being too polar. Despite a higher retention provided by PRP-1, the retention of the hydroxy-dealkylated derivatives is too low to allow the handling of a sufficiently large volume for determinations below the μg/l level. The more polar derivatives, i.e., ammeline, ammelide and cyanuric acid, are not retained by either C18 silica or PRP-1, whereas the capacity factor measured on a PGC column eluted with water was above 100. The conditions of extraction and elution using a cartridge packed with 500 mg of PGC are described for all the series of compounds and recoveries above 90% were obtained from 500-ml samples. The analytical separation was carried out using a C18 silica column and confirmation was obtained using a PGC column. Determinations of the degradation products in real samples are presented with detection limits below the 0.1 μg/l level.