[Low Temperature Properties of T Shape Photoacoustic Cell and Applications].

Niu Ming-sheng,Liu Qiang,Wang Gui-shi,Yuan Yi-qian,Huang Wei,Zhang Wei-jun,Gao Xiao-ming
DOI: https://doi.org/10.3964/j.issn.1000-0593(2013)03-0577-05
2013-01-01
Abstract:A homemade low-temperature T shape photoacoustic system (PAS) in the range from 0 to - 100 degrees C is well established. The mode distribution, the temperature dependence of the resonant frequency, Q-factor and the sensitivity of the PAS were investigated. The measurement of the carbon dioxide in the range from 0 to -100 degrees C showed that the system can be used in trace gas detection under low-temperature. The results show that the PAS can be used to study the atmospheric molecules continuous absorption at low-temperature.
What problem does this paper attempt to address?