The Research on Detector Sensitivity of Full- Coverage Animal PET System

YANG Qun,XIA Yan,MA Tian-yu,LIU Ya-qiang,WANG Shi
DOI: https://doi.org/10.3969/j.issn.0258-0934.2012.03.003
2012-01-01
Abstract:In order to improve detector sensitivity of small animal PET system and increase geometry angle coverage rate of detector,a full-coverage animal PET system design is proposed,in which two square detectors are added to the original circular PET scanner's both ends.There is a hole in the middle of each of the two square detectors.To investigate the detector sensitivity performance of the new system,GATE was used to build system model and simulation was performed with three different types of radioactive sources which is point,plane and volume sources.The simulation results demonstrate that the change to the detector structure effectively increase the detector sensitivity,and the number of line of responses(LOR) is doubled more than before which will potentially benefit image reconstruction;meanwhile the simulation results of point and plane radioactive sources show that the proposed full-coverage arrangement of detectors improves the uniformity of sensitivity in the FOV,however the hole on the board detector weakens the improvement of detector sensitivity,especially to the radioactive source which is put on the edge of the FOV.Full-coverage animal PET system as proposed can improve detector sensitivity and image quality.