Effect of an iliac bone plate on the subsidence of iliac screw after fatigue loading

YU Binsheng,ZENG Liwen,LI Zemin
2011-01-01
Abstract:Objective:To evaluate the biomechanical effect of a self-designed iliac bone plate on the subsidence of iliac screw after fatigue loading.Method:Lumbo-pelvic specimens from 6 adult fresh cadavers were used in this study,their mean bone mineral densities(BMD) of L1-L4 were measured by using dual-energy radiograph absorptiometry.The iliac bone plate was randomly placed at the iliac screw entry point of one-side ilium(group A),while the other side without iliac bone plate served as the control(group B).Then,the iliac screws with 70mm length and 7.5mm diameter were inserted into the ilium bilaterally.On a MTS testing machine,a dynamic compressive loading of 100-400N was added to the screw head for 20000 cycles.The screw subsidence displacements following each 5000 cyclic loadings were measured and compared between two groups.Result:The average BMD value of 6 specimens was 0.89±0.06g/cm2,ranged from 0.78g/cm2 to 1.06g/cm2.All the screws passed the 20000 cyclic loadings smoothly.At the time-points of the 5000th,10000th,15000th,and 20000th cyclic loadings,the subsidence displacements of iliac screw in group A were 0.28mm,0.36mm,0.48mm and 0.66mm respectively;while those in group B were 0.34mm,0.69mm,0.88mm and 1.07mm respectively.After the first 5000 cyclic loadings,no significance with respect to subsidence displacement was detected between two groups(P>0.05).However,at the 10000th,15000th and 20000th cycles,significant difference was noted between two groups(P<0.05).After the 20000 cyclic loadings,no screw in group A and 5 screws in group B were found loosen under radiograph.Conclusion:The iliac bone plate can effectively resist the subsidence of iliac screw following fatigue loading and protect iliac screw against loosening.
What problem does this paper attempt to address?