Chemiluminescence Microflow Injection Analysis System on a Chip for the Determination of Uric Acid Without Enzyme.

DY He,ZJ Zhang,Y Huang,YF Hu,HJ Zhou,DL Chen
DOI: https://doi.org/10.1002/bio.847
IF: 2.613
2005-01-01
Luminescence
Abstract:A new microflow injection analysis (microFIA) system on a chip coupled with chemiluminescence (CL) for the non-enzymatic determination of uric acid is described. The microFIA system produced by using two transparent poly(methylmethacrylate) (PMMA) chips measured 50 x 40 x 5 mm, the microchannels, etched by CO2 laser, were 200 microm wide and 100 microm deep, and the volume of the reaction area (RA) was about 1.2 microL. The injection pump, with accurate time control, monitored all reagents, including the sample. The uric acid was sensed by the chemiluminescence reaction between luminol and ferricyanide. The linear range of the uric acid concentration was 0.8-30 mg/L and the detection limit was 0.5 mg/L (S/N = 3). The relative standard deviation was 4.42% for 5 mg/L uric acid (n = 8). The proposed method has been successfully applied to the non-separation determination of uric acid in human serum and urine.
What problem does this paper attempt to address?