Evaluation of 16-Slice Spiral CT Coronary Angiography in Diagnosis Coronary Artery Stenosis

许尚文,成官迅,陈自谦,张雪林,曾建华,李天然,钱根年
DOI: https://doi.org/10.3969/j.issn.1671-5098.2006.21.002
2006-01-01
Abstract:Objective To evaluate the clinical value of 16-slices spiral CT coronary angiography in the detection of coronary artery stenosis.Methods 46 patients were detected with ECG-gated MSCT.And all patients were also investigated with SCA.Results of the determination of≥50% stenoses of the vessel diameter with MSCT were compared to findings of SCA.Results The sensitivity,specificity,accuracy,positive and negative prediction value of MSCTCA were 82.22%、95.68%、92.39%、86.05% and 94.33%,respectively.Conclusion 16-slice spiral CT coronary angiography is a noninvasive,simple and safe method.16-slice spiral CT coronary angiography has a good clinical value in the diagnosing of ≥50% stenoses of the vessel diameter.
What problem does this paper attempt to address?