Diagnostic Value of 16-Slice Spiral Computed Tomography for Detecting Coronary Artery Disease

毛定飚,滑炎卿,张国桢,王鸣鹏,丁其勇,葛虓俊
DOI: https://doi.org/10.3969/j.issn.1673-6583.2004.01.018
2004-01-01
Abstract:Objective:To evaluate the accuracy and the limit of 16-slice spiral CT in diagnosis of coronary artery disease. Methods: 20 patients were studied by 16-slice spiral CT (retrospectively ECG-gating; 0. 42s rotation). The results were compared with coronary artery angiography. Results: In the 136 coronary artery segments (vessel diameter≥2mm)of 20 patients,122 were judged to be evaluable by CT(89. 7%). The sensitivity and the specificity for the detection of stenosis≥50% were 87. 5 % and 96. 6% respectively. When 14 coronary artery segments were judged to be unevaluable were included in the analysis, the sensitivity was 77. 7%. Conclusions: 16-slice spiral CT could be used in noninvasive assessment of significant coronary artery stenosis with high accuracy. The limit was inadequate temporal resolution and unavoidable artifacts.
What problem does this paper attempt to address?