Evaluation of coronary bypass grafts with multislice spiral CT angiography

GAO Li,WANG Ji-chen,QIU Jian-xing,SUN Xiao-wei,GAO Fu-sheng,TANG Guang-jian,JIANG Xue-xiang
DOI: https://doi.org/10.3321/j.issn:1003-3289.2006.10.013
2006-01-01
Abstract:Objective To evaluate the value of multislice CT angiography (CTA) for assessment of coronary artery bypass graft (CABG). Methods A total of 28 patients with 69 coronary artery bypass grafts were examined by CTA with 16-slice and 64-slice CT, in which 26 coronary artery bypass grafts of 10 patients coronary artery bypass grafts were examined by coronary angiography as well. Using multiple linear regression to evaluate the effect of patient's heart rate, fluctuation of heart rate, breathing movement and different CT scanners on image quality of CT coronary angiogram. Results The patient's breathing movement was negatively related to the image quality, the value of standardized coefficients was -0.838. The image quality of 64-slice CT was higher than that of 16-slice CT. Taken the CAG as reference standard, the sensitivity and specificity of CT in evaluating the anastomosis and bypass grafts were 95.7% and 92.3% respectively. Conclusion MSCT can provide high image quality of CABG, which has great potential value in the follow-up of this disease.
What problem does this paper attempt to address?