Determination of Warfarin in Human Blood Serum by RP-HPLC

CHEN Yun,HUANG Xiao-hui,TANG Hai-qin,YANG Wei-wei,ZHU Yan-wu,JIN Yong,LI Jun
DOI: https://doi.org/10.3969/j.issn.1009-6469.2007.12.004
2007-01-01
Abstract:Aim To establish an method for determination of warfarin in human's blood serum.Methods RP-HPLC method was used,with C18 column at the standard of 150 mm×4.6 mm,0.01 mol·L-1 KH2PO4 methanol-H3PO4 solution(38∶62∶0.009) as mobile phase.The flow rate was 1.0 ml·min-1.UV detection was determined at 205 nm.Results The reserve time of Warfarin in human blood serum was 8.7 min.The linearity range of standard curve was between 0.0625 and 4.0 mg·L-1.The average recovery was 90%.RSD of three consistencies including the high,middle and low varied in less than 10%.Conclusion The method is simple,fast and accurate for the analysis of warfarin in human blood serum.