Determination of Bivalirudin in Human Blood by HPLC-MS/MS

LI Jing,LIN Yang,WU Wei,ZHAO Gui-ping,SUO Wei,DU Hai-yan,QIU Qi
DOI: https://doi.org/10.11669/cpj.2013.16.017
2013-01-01
Abstract:OBJECTIVE To establish an HPLC-MS /MS method for the determination of blood-bivalirudin concentration.METHODS Following a SPE method,the analytes were separated on a C 18 column interfaced with a triple-quadrupole tandem mass spectrometer using positive electrospray ionization.RESULTS The calibration linear rang of blood-bivalirudin concentration was 25-20 000 μg·L-1(r = 0.996 9).The between-day and within-day RSD% were less than 6.63%.The recoveries were more than 80%.CONCLUSION The rapid,sensitive and reliable HPLC-MS / MS method can be applied to monitor blood-bivalirudin profile.
What problem does this paper attempt to address?