Abstract:The direct determination of trace Se and Te in high purity silver by inductively coupled plasma mass spectrometer was studied.The optimum instrument parameters and work conditions were selected.The effects of silver matrix on determination of Se and Te were eliminated by adding Cs as internal standard and the determination results were accurate and reliable.The determination limits for Se and Te were 0.1 and 0.06μg/g and the RSD were less than 5% with 92.6%~103.5% recoveries.The method has been applied to the determination of Se and Te in high purity silver with satisfactory results.