Preparation of Ph Microelectrode and Its Application in Plaque Ph Measurement

XU Xiao-juan,LIU Lu-chuan,YUAN Ruo,ZHAO Qin
DOI: https://doi.org/10.3969/j.issn.1003-9872.2006.04.017
2006-01-01
Abstract:Objective To develop a chemically modified pH microelectrode which can be used in the measurement of dental plaque pH in situ. Methods Iridium oxide (IrO2) was used as the H+ sensitive membrane and attached to the surface of platinum wires by electroplating. Then,the pH responses of the microelectrodes were tested and the better ones were chosen for the measurement of the change of dental plaque pH within 1h after sucrose rinsing. The plaque of the symmetrical site was measured by the Beetrode pH microelectrode as a contrast. Results The homemade microelectrodes showed a linear response in the pH ranging from 1 to 11with response sensitivity of 55.8 mV/pH. The response time was about 15~20 s and potential drift was 0.5~1.0 mV/pH within 8h. Its life-span could be maintained at least 1 year. In the measurement of plaque pH,there were no statistical differences between the homemade and the Beetrode microelectrode. Conclusion The homemade IrO_2 microelectrodes show very promising pH sensing performance. They are considered to be feasible in the in situ measurement of plaque pH because of the good stability over a wide pH range,rapid response and being tiny enough to be inserted into the interproximal site of the teeth.
What problem does this paper attempt to address?