Analysis of P-Xylene Oxidation Products by GC and HPLC

XU Hai-bo,WANG Qin-bo,LI Xi
DOI: https://doi.org/10.3969/j.issn.1008-8261.2006.02.006
2006-01-01
Abstract:A method that use gas chromatography(GC) to rapid quantify the solvent acetic acid and use high performance liquid chromatography(HPLC)to rapid analyse all components contain benzene in PX oxidation system was provided.By using Agilent 1100 HPLC and Diamonsilc 18((D5 μm×4.6 mm×150 mm)chromatography column,using methanol-acetonitrile-water as mobile phase,using external standard method three-component grade elution program,the 4-CBA and TA content in solid state product was analysed.By using isopropyl benzene as the internal standard material,using Agilent 1100 HPLC,using methanol-acetonitrile-water as mobile phase,using three-component grade elution program,the components that contain benzene such as TALD,PT,4-CBA,TA,BA,PX and internal standard material isopropyl benzene in liquid phase product was analysed.Shimazu6A GC was used to analyse the solvent acetic acid and internal standard material isopropyl benzne in liquid phase.The gas phase and liquid phase results are interrelated through internal standard material isopropyl benzne.The method is simple and rapid.The analysis period is short.