Quality analysis on medical laboratory pure water
刘晓婷,向代军,邵安良,王成彬
DOI: https://doi.org/10.3969/J.ISSN.1672-8270.2016.02.033
2016-01-01
Abstract:Objective:To analysis pure water quality of medical laboratory before and after purification.Methods: According to GB/T 6682-2008: water for analytical laboratory use: specifications and test methods, explore pure water quality before and after purification from pH, conductivity, oxidized substances, absorbance, evaporation residue and the content of soluble silicon, respectively.Results: Before purification, pH value, conductivity, oxidized substances, absorbance, evaporation residue and the content of soluble silicon of pure water was 7.4, 0.07 mS/m, 0.07 mg/L, 0.008, 0.8 mg/L and 0.015 mg/L, respectively. And it conformed to 2nd grade. While pH value, conductivity, oxidized substances, absorbance, evaporation residue and the content of soluble silicon of pure water is 7.18, 0.006 mS/m, 0.06 mg/L, 0.001, 0.6 mg/L and 0.01 mg/L, respectively after purification, which was in line with 1st grade.Conclusion: Pure water quality of medical laboratory is 2nd grade before purification, while water quality increases to 1st grade after purification. Purification has been improved pure water quality significantly, and it can meet the requirements of medical laboratory.