Experiment Study of Gas Rotameter's Measuring Error

PIAO Li-hua,ZHANG Tao,SUN Li-jun,BI Li-hong
DOI: https://doi.org/10.3969/j.issn.1000-3932.2010.03.018
2010-01-01
Abstract:The gas float flowmeter was firstly graduated in the water flow standard facility.Then the measurement errors were studied in water flow standard facility and gas flow standard facility.It was founded that the measurement values of gas float flowmeter tested in the water facility were greater than true values.The reason was the fluid dynamic viscosity for calibration was higher than that of fluid for measurement.So in order to ensure the accuracy of gas float flowmeter,the impact of fluid viscosity must be taken into account if the gas float flowmeter was calibrated in the liquid flow standard facility.And the viscosity coefficient k should be introduced in the scale conversion formula.
What problem does this paper attempt to address?