Analysis of Measurement Errors of Coriolis Mass Flowmeter with Straight Cylindrical Tube

Bin Li
2014-01-01
Abstract:The theory of Coriolis mass flowmeter with straight cylindrical tube was presented. The experimental instrument and methodology were introduced. The mass flows obtained from experiments were close to actual values,with some errors. The causes and the characteristics of the errors are analyzed by studying many factors,such as the experimental devices,geometric characteristics,and vibration parameters. The excitation forces and the excitation frequencies can affect the measurement errors according to the analysis by theories and experiments. Furthermore,there are some non- negligible errors between the actual values and the experimental results when experiments are performed in the natural frequencies. These frequencies should be avoided during the measurement.
What problem does this paper attempt to address?