Comparisons between thermodilution method and arterial pressure cardiac output with rapid volume infusion

FAN Xiao-hua,LIU Ji,FAN Qin,BAO Rui,LI Bo,DENG Xiao-ming
2012-01-01
Abstract:Objective To compare the thermodilution method with the arterial pressure when cardiac output (CO)is weasured, and observe the hemodynamic changes with rapid infusion.Methods After the anesthesia induction of 10 patients (ASA Ⅰ or Ⅱ) with sacral tumors, Swan-Ganz catheter, Flotrac sensor and Vigileo monitor were placed. Intermittent cardiac output (ICO) was measured discontinuously transpulmonary thremodilution and arterial pressure cordiac output(APCO) was recorded simultaneously. A dose of 10 ml/kg of hydroxyethyl starch 130/0.4 was infused within 30 minutes before the operation. HR, MAP, CVP, arterial wedge pressure (PAWP), ICO stroke volume variability (SVV), APCO and stroke volume index (SVI) were measured before and after rapid infusion. Bland-Altman analysis was used to compare the difference of CO measured by the two methods.Results The mean of difference between APCO and ICO was 0.10 L/min, and 95%CI interval was -1.14-0.94 L/min. The mean of average (APCO plus ICO) was 5.97 L/min and could be accepted while considering the maximum absolute of 95% interval of 1.14 L/min. After rapid infusion, CVP, PAWP, APCO and SVI increased (P<0.05 or P<0.01), SVV decreased (P<0.01) while HR and MAP had no significant changes.Conclusion APCO is well related to thermodilution method and may provide effective perioperative hemodynamic monitoring.