Nanometer-Sized ZrO_2 Microcolumn Separation Combined with ICP-MS for the Speciation of Chromium in Water Samples

曾艳,占红梅,胡斌
DOI: https://doi.org/10.3969/j.issn.1006-6144.2010.02.001
2010-01-01
Abstract:A simple and convenient method has been developed for the speciation of Cr(Ⅲ) and Cr(Ⅵ) in water samples by nanometer-sized ZrO2 microcolumn separation and ICP-MS detection. The adsorption behaviors of Cr(Ⅲ) and Cr(Ⅵ) on nanometer-sized ZrO2 were studied and it was found that at pH=8.0 Cr(Ⅲ) could be quantitatively adsorbed by the microcolumn and Cr(Ⅵ) passed through the microcolumn directly. The main factors affecting the separation of Cr(Ⅲ) and Cr(Ⅵ) on nanometer-sized ZrO2 were studied. Under the optimal conditions,the relative standard deviation for chromium was 2.2%(c=100 ng/mL,n=9),the limit of detection(3σ)was 0.06 ng/mL and the limit of determination (10σ) was 0.37 ng/mL. No obvious interference has been observed from the co-existed species. The proposed method has been successfully applied to the speciation of chromium in real water samples.