Radiation dosage accepted by children during interventional treatment for congenital heart disease

FENG Jun,CHENG Jing-lin,GUO Jie
DOI: https://doi.org/10.3969/j.issn.1009-6469.2012.02.018
2012-01-01
Abstract:Objective To estimate the radiation dose to which children are exposed during cardiac catheterizations for the treatment of congenital heart disease to analyze the factors affecting the radiation dose and to find out the measures to decrease the radiation dose.Methods 91 cases undergoing cardiovascular interventional procedures were collected from January to July 2011 in our hospital,including 36 atrial septal defect(ASD),25 patent ductus arteriosus(PDA),30 ventricular septal defect(VSD) to observe cumulative dose(CD),dose area product(DAP),fluoroscopy time and those differences.Results The cumulative dose of the three groups of children were(0.15±0.10),(0.31±0.29),(0.37±0.18) Gy,and DAP were(15.78±10.50),(33.11±31.35),(37.69±21.56) Gy·cm2,and fluoroscopy time was(8.64±5.10),(8.15±5.23),(10.47±7.18)min,respectively.CD values,DAP values and fluoroscopy time were quite different between different operators.Conclusion Intervention therapy for congenital heart disease,including patent ductus arteriosus,atrial septal defect,and ventricular septal defects is safe,effective and reliable,but children are more sensitive to radiation.Operators should improve skills and shortern radiation time to reduce dose.
What problem does this paper attempt to address?