Abstract:[Objective]To test the suitability of N95 particulate matter protection mask of 3MTM1860 type for medical use,and promote the awareness of wearing and adjusting method of it among medical staff.[Methods]The suitability of N95 particulate matter protection mask of 3MTM1860 type for medical staff were tested with qualitative method by 3MTMFT-30 qualitative suitability test equipment.[Results]Of 637 medical workers,apart from 13(2.04%) were insensitive to bitter,the others participated in the test were classified into 556(87.28%) of level 1,49(7.69%) of level 2 and 19(2.98%) of level 3,according to the sensitivity;611(95.92%) passed qualitative suitability test,but 26(4.08%) did not.[Conclusion]Qualitative suitability test not only lets the medical workers understand the correct wearing method of mask,but also facilitate their choice of appropriate mask based on individual factors such as facial form.