Influence of scatter radius parameter change on the results of intensity modulated radiation therapy inverse optimization in CadPlan/Helios planning system

ZOU Li-fen,FU Xiao-long,CHENG Qing-fang,WANG Sheng-zi,WANG Li-juan,邹丽芬,傅小龙,程庆芳,王胜资,王丽娟
DOI: https://doi.org/10.3760/j.issn:1004-4221.2006.02.015
2006-01-01
Abstract:Objective To investigate the influence of scatter radius parameter change on the results of intensity modulated radiation therapy (IMRT) inverse optimization in CadPlan/Helios planning system. Methods Using a homogeneity water equivalent rectangular phantom, the dose gradients at the interface of target volume and normal tissue was evaluated under different scatter radii. The dose distribution of a nasopharyngeal carcinoma patient with 7 field IMRT after inverse optimization with different scatter radii parameters was compared. Results s The dose gradient increased with larger scatter radii of inverse planning optimization in both target and normal tissue. With dynamic Multi-leaf collimator (MLC) deliver, the dose gradient could be even higher than that of the principal jaws. When the same case was optimized with different scatter radii, the difference of DVH between real-time and three-dimension was more significant than that of shorter scatter radii. The target dose was close to prescription dose and the normal tissue was protected much better when using larger scatter radii. Conclusion Larger scatter radii should be used in inverse planning optimization.