Abstract:Objective:To evaluate the threshold of parameter on bacterial particle of UF-100.Method:Urine quantitative cultures of 462 samples from hospitalized patients were used as a reference method for comparison.Threshold of parameter on bacterial particle was determined by ROC curve.The association between the quantity of bacterial particle measured by UF-100 and the quantity of urine cultures was demostrated by regression curve.Results:The threshold of bacterial particle were 5 459/μl,6 520/μl and 8 092/μl compared with culture at a cutoff of 102 CFU/ml,104 CFU/ml and 105 CFU/ml,respectively.The threshold of bacterial particle were influenced by gender and species of pathogen. The regression curves of the quantity(X) of bacterial particle and the logarithm(Y) of the quantity of cultures were Y=1.930 2+0.000 2X and Y=-6.141 2+1.121 4×lnX.Conclusions:The threshold of bacterial particle is determined by urine quantitative culture and influenced by gender and species of pathogen.The results of urine culture can be estimated by the regression curve.