Determination of the mass fraction of Ni in fluid cracking catalyst by X - ray fluorescence spectrometry

LIU Ming-xia,LIU Rui-qiang,SUN Wei,WANG Ya-hong,GAO Xiong-hou,ZHANG Zhong-dong,ZHANG Hai-tao
DOI: https://doi.org/10.3969/j.issn.1009-0045.2013.03.017
2013-01-01
Abstract:X-ray fluorescence spectrometry for determining the mass fraction of Ni in fluid cracking catalyst was established with Ni containing standard samples made by impregnating method,and the working curve obtained by the method was corrected with intensity correction model.The results showed that the accuracy of the method was affected by rare earth elements.Using X-ray fluorescence spectrometry with matrix correction,the relative errors of the mass fraction of Ni between the measured and theoretical values were within ± 3%,the relative standard deviations were lower than 0.300%,implying this method had higher accuracy and precision.