Application of X-ray fluorescence spectrometry in catalyst composition determination without standard

LIU Yan,SUN Yang,LI Dong-mei,SU Dan,SHEN Qi-hui,YU Dong-dong
DOI: https://doi.org/10.13595/j.cnki.issn1000-0720.2017.0299
2017-01-01
Abstract:A method was established for the determination of the major and minor elements content in the catalyst by the method of X-ray fluorescence spectrometry without sample analysis in this paper.Firstly,the powder catalyst will be molten into glass samples.Then metal content will be tested,such as aluminum,silicon,zirconium,cerium,and lanthanum elements by the software UniQuant without standard sample analysis.The determination results of relative standard deviation (RSD,n =9) was from 0.11% to 1.9%.Finally the results were coincided with the chemical analysis method.