A Spectrophotometric Procedure for DNA Assay with a Micro-Sequential Injection Lab-on-valve Meso-Fluidic System

XW Chen,JH Wang,ZL Fang
DOI: https://doi.org/10.1016/j.talanta.2005.02.022
IF: 6.1
2005-01-01
Talanta
Abstract:A micro-sequential injection spectrophotometric procedure for DNA assay was developed based on the employment of a lab-on-valve (LOV) meso-fluidic analytical system. A small amount of crystal violet solution (10mul) was de-colored inside the flow cell of the LOV at the presence of 5mul lambda-DNA/HindIII within a certain pH range, and the absorbance decrease of crystal violet solution at 591nm was measured via optical fibers and was employed as the basis of quantification. A uni-variant approach was adopted for the optimization of experimental parameters, including buffer pH, concentration and volume of crystal violet solution, reaction time and sample/reagent loading flow rates. A linear calibration graph was obtained within 0.2-6.0mugml(-1), along with a detection limit of 0.07mugml(-1). The procedure was applied for the determination of lambda-DNA/HindIII in synthetic samples in comparison with a documented procedure.
What problem does this paper attempt to address?