Microfluidic Chip Detection Instrument With High Signal-To-Noise Ratio
El Cai,Zf Cen,Xt Li,Yq Lin,Xm Huang
2004-01-01
Abstract:A Microfluidic Chip Detection instrument must have a structure with high sensitivity, high signal-to-noise ratio and reaction speed, because that Microfluidic Chips have been designed microchannel construction, also that the instrument has been required to detect extremely low density reagent. The instrument applies Laser Induced Fluorescence(LIF) principle, adopt the confocal theory in detect optical pass. Moreover, the instrument uses electron-multiplier phototube coherent electric circuit (Laser outlet wavelength 473 nm, fluorescent reagent FITC). Furthermore, the instrument has disposed the possible stray light by the way of using special wavelength filter, choosing pinhole with proper diameter, doing some special treatment with all the apartments. On the other hand, six-way high voltage for single controlling and well man-machine interface has also been designed. At the same time, for better SNF, we have also programmed relevant software and designed some useful algorithm. This instrument's low detecting limit can reach 1 nmol/L.