Microflow injection potassium bioassay based on G-quadruplex DNAzyme-enhanced chemiluminescence.

Lifang Song,Xiaoyan Pan,Hong Shen,Yaling Yu
DOI: https://doi.org/10.1002/bio.2661
IF: 2.613
2014-01-01
Luminescence
Abstract:By taking advantage of microflow injection chemiluminescence analysis, we developed a distinctive microfluidic bioassay method based on G-Quadruplex DNAzyme-enhanced chemiluminescence for the determination of K+ in human serum. AGRO100, the G-rich oligonucleotide with high hemin binding affinity was primarily selected as a K+ recognition element. In the presence of K+, AGRO100 folded into G-quadruplex and bound hemin to form DNAzyme, which catalyzed the oxidation of luminol by H2O2 to produce chemiluminescence. The intensity of chemiluminescence increased with the K+ concentration. In the study, the DNAzyme showed both long-term stability and high catalytic activity; other common cations at their physiological concentration did not cause notable interference. With only 6.7x10(-13)mol of AGRO100 consumption per sample, a linear response of K+ ranged from 1 to 300 mu mol/L, the concentration detection limit 0.69 mu mol/L (S/N=3) and the absolute detection limit 1.38x10(-12)mol were obtained. The precision of 10 replicate measurements of 60 mu mol/LK+ was found to be 1.72% (relative standard deviation). The accuracy of the method was demonstrated by analyzing real human serum samples. Copyright (c) 2014 John Wiley & Sons, Ltd.
What problem does this paper attempt to address?