Observing Affection That CK and CK-MB Reagents Made in Measuring TG with Biochemical Automatic Analysis

GUO Qun,HU Jin-cao,GU Guang-yu
DOI: https://doi.org/10.3969/j.issn.1671-7414.2007.01.023
2007-01-01
Abstract:Objective To research the items and solution tothe crossed contamination which the triglycerides(TG) automatic analysis make.Methods Added the reagents,which was suspicious of crossed contamination to the triglycerides detection,to the pooled serum as experiment tube to measure the triglycerides,compared to the comparison tube to which the same proportion normal saline was added,primarily filtering the items of crossed contamination to the triglycerides detection.Firstly,Measured the creatine kinase(CK) and the creatine kinase isozyme(CK-MB),the items of crossed contamination to the triglycerides detection in the primarily filtering experiment,then measured the TG for 9 times continuously,observing the influence on the results of TG measuring that automatical analysis made,at the same time,observing the continuous extent to the influence on the results of TG measuring that the CK andCK-MB reagents made.Results CK and CK-MB reagents had obvious crossed contamination to the results of the triglycerides measuring.The analytical sequence that measuring CK firstly and then CK-MB had obvious influence on the results of the triglycerides measuring.The influence which the CK andCK-MB reagents made can continued the second tube.Conclusion CK and CK-MB reagents have obvious crossed contamination to the results of the triglycerides measuring.The analytical sequence has the influence on the experimental results that must be paid attention in the biochemical automatic analysis.
What problem does this paper attempt to address?