A Rare Case of Falsely Elevated High-Sensitivity Cardiac Troponin T Due to Antibody Interference and a Literature Review

Cheng Fang,Jiawen Chen,Ben‐Lai Wu,Xing Yang,Huang Yuan
DOI: https://doi.org/10.7754/clin.lab.2023.230230
IF: 1.053
2023-01-01
Clinical Laboratory
Abstract:This study aimed to demonstrate potential measures to discern interference and establish an algorithm for identifying interference in an hs-cTnT assay with a clinical discrepancy.Laboratory error and disease factors were first excluded. Then, a serial dilution test, polyethylene glycol (PEG) precipitation, and heterophile antibody blocking reagents (HBRs) were performed to confirm the interference substance. In addition, high-speed centrifugation and western blot were tried.The sample diluted linearly (R2 = 0.9966). However, there was a dramatic reduction in the concentration of hs-cTnT concentration to a normal level after both PEG precipitation and HBR1 treatment. The results were inconclusive in high-speed centrifugation and western blot assay.This study elucidated a heterophile antibody interference in measurement of hs-cTnT alone on a Roche analyzer.
What problem does this paper attempt to address?