An Unusual Case of False Positive

Michela Salvatici,Monica Gaimarri,Francesca Rispoli,Barbara Bianchi,Delia Francesca Sansico,Eleonora Matteucci,Andrea Antonelli,Francesco Bandera,Lorenzo Drago
DOI: https://doi.org/10.3390/ijms252211937
IF: 5.6
2024-11-07
International Journal of Molecular Sciences
Abstract:Heterophile antibodies, which can arise from infections, autoimmune disorders, or exposure to animal antigens, can interfere with immunoassays. These antibodies can cross-react with the test reagents used in troponin assays, causing a false elevation in troponin levels. The paper describes a case of a 37-year-old male drug abuser admitted to the emergency room with chest pain. A series of troponin measurements performed using different assays gave discrepant results. Only thanks to the use of Scantibodies HBT tubes, which remove heterophile antibodies, was it possible to make a correct diagnosis of troponin negativity. In conclusion, a correct laboratory/clinical approach to the identification of heterophile antibody interference is essential for accurate troponin testing in order to avoid false positive results. Implementing neutralizing tests can significantly improve the reliability of these diagnostic assays, ensuring better patient outcome.
biochemistry & molecular biology,chemistry, multidisciplinary
What problem does this paper attempt to address?