The influence of M protein interference on the routine biochemical assays of serum samples

SHEN Minna,WU Jiong,GUO Wei,ZHANG Chunyan,SONG Binbin,WANG Beili,WANG Yang,PAN Baishen
DOI: https://doi.org/10.3969/j.issn.1673-8640.2011.11.004
IF: 1.865
2011-01-01
Laboratory Medicine
Abstract:Objective To investigate the influence and potential interference of monoclonal protein(M protein) on 24 routine biochemical assays,and establish the operational processes for clinical laboratory to identify,confirm and remove the interference of M protein.Methods The reaction absorbance curves of 57 M protein positive serum samples were observed.According to the reaction absorbance curves,the suspected samples interfered by M protein were identified.Furthermore,the interference of M protein was confirmed by dry chemical method(Johnson Vitros 5.1 FS automatic biochemical analyzer).The samples interfered by M protein were diluted by normal saline to remove the interference.The high-density lipoprotein cholesterol(HDL-C) assay was performed manually to explore the causes of M protein interference.Results A total of 50(87.7%) of 57 samples(M protein concentration: 2.0-71.4 g/L) did not show any interference.M protein interference occurred not obviously in 7(12.3%) samples,and there were M protein interferences in 2 of the 24 routine biochemical assays.The levels of C reactive protein(CRP) decreased falsely in 6 samples,and there was negative result of HDL-C in 1 sample.The false decreasing of CRP caused by M protein interference was correlated with CRP concentration rather than M protein concentration.In other words,the higher the actual value of CRP was,the severer the M protein negatively interfered.The frequency of the M protein interference was correlated with the concentration and type of M protein.In other words,the higher the M protein concentration was,the higher the frequency of the interference occurred.M protein interference was commonly caused by IgG λ and κ.After diluting the interfered samples,the M protein interference can be removed.Conclusions In the 24 routine biochemical assays,M protein interference occurs relatively in CRP and HDL-C assays.M protein interference can be removed via diluting its concentration or retesting with dry chemical method to obtain the reliable results.
What problem does this paper attempt to address?