Abstract:A gas chromatographic method for determination of organic solvents residual in CHU-9126 was established. A DB-624 capillary gas chromatographic column coated with 94% dimethylpolysiloxane and 6% cyanopropylpolysiloxane stationary phase was used with the carrier gas of nitrogen at a flow rate of 2ml·min~ -1 . The initial column temperature was kept at 70℃for 3 minutes, then raised to 90℃ at the rate of 3℃·min~ -1 and kept for 7 minutes, then raised to 220℃ at the rate of 20℃·min~ -1 . There were good relationship between the area and amounts of ethanol, tetrahydrofuran and N,N-dimethylformamide. The average recovery rates were 99.9%, 100.3% and 99.6%, and limit of detection were 0.12, 0.10 and 0.28ng respectively. This method appeared to be sensitive, accurate, repeatable and easy to operate.