Development of an amperometric glucose biosensor based on the immobilization of glucose oxidase in an ormosil-PVA matrix onto a Prussian Blue modified electrode

HaiLing Chen,Li Zhao,Xi Chen,ZhiXia Zhuang,XiaoRu Wang
DOI: https://doi.org/10.1007/s11426-009-0146-4
2009-01-01
Abstract:An amperometric glucose biosensor was developed based on the immobilization of glucose oxidase in the organically modified silicate (ormosil)-polyvinyl acetate (PVA) matrix onto a Prussian Blue (PB)-modified glassy carbon electrode. A higher stability PB-modified electrode was prepared by the electrochemical deposition of FeCl 3 , K 3 [Fe(CN) 6 ] and ethylenediamine tetraacetic acid (EDTA) under cyclic voltammetric (CV) conditions. The effects of the potential range of CV conditions, electrolyte cations, applied potential, pH, temperature and co-existing substances were investigated. The detection limit of the glucose biosensor was 8.1 μmol·L −1 (S/N = 3) with a linear range from 20 μmol·L −1 to 2 mmol·L −1 ( R = 0.9965). The biosensor presented a fast response and good selectivity. Additionally, excellent reproducibility and stability of the biosensor were observed.
What problem does this paper attempt to address?