EVALUATION OF UNCERTAINTY IN MEASUREMENT OF DETECTOR SENSITIVITY AND DETECTION LIMIT FOR GAS CHROMATOGRAPH

Lü Hong,Liu Zhijuan,Xu Feng,Zhang Wenshen,Xun Qining
DOI: https://doi.org/10.3969/j.issn.1008-6145.2010.01.004
2010-01-01
Abstract:The uncertainty of main technical parameters sensitivity and detection limit of the detector in verification of gas chromatograph were evaluated according to JJF 1059-1999 Evaluation and Expression of Uncertainty in Measurement and JJG 700-1999 Examination Instruction of Gas Chromatograph. The components of uncertainty were analyzed. The mathematical model of measurement was confirmed,and the expanded uncertainty of measurement result was calculated. The relative expanded uncertainty of sensitivity or detection limit was 4.4 (k=2) for concentration dependent detector,and was 3.8 (k=2) for mass detector. The rationality of evaluation result was validated contrastingly according to JJF 1033-2008.